EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 81
NPD Response Low
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
Perform column maintenance: Bake out contaminants as
needed. Verify correct column installation.
A large concentration of solvent has extinguished the
hydrogen/air plasma. Increase the bead voltage. Run the
makeup gas at a flow rate of 5 mL/min.
Verify that there is hydrogen coming from the external
supply. Check that flow and pressure are turned on at the
keyboard. The hydrogen flow rate should be between 1.0 and
5.5 mL/min. Measure the actual gas flow at the detector. (See
“To Measure a Detector Flow”.)
Check for a partially plugged jet. See To Check for a Plugged
FID Jet.
If the upper ceramic insulator is contaminated, a high offset
(2 to 15 pA or more) will occur when the bead is off. This
directly affects sensitivity. Replace the ceramic insulator.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals