EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
104 Agilent 7890 Series Troubleshooting
4 GC Not Ready Symptoms
FID or NPD Measured Hydrogen and Makeup Gas Flows Much Less Than
Setpoint
Check for a clogged or partially clogged jet. A clogged jet
creates backpressure. Since the flow module uses pressure
control, the increased backpressure simulates proper flow.
The actual flow rate will drop but the GC remains functional.
See:
To Check for a Plugged FID Jet
To Check for a Plugged NPD Jet
Check for a column fitting leak at the detector base.
Replace the FID or NPD jet.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals