Checking for Leaks 7
Agilent 7890 Series Troubleshooting 171
To Correct Leaks in the Volatiles Interface
If the inlet fails a pressure decay test, check the following:
The caps and plugs used in the test—make sure each is
correctly installed and tight.
If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
The split vent and pressure sensing connections at the
interface.
The sampler connection to the interface.
The sampler.
If these items do not resolve the problem, contact Agilent for
service.