Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 77
FPD Low Peak Areas
Check the flow settings, then measure the actual detector
flows. (See “To Measure a Detector Flow”.)
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
Verify that the column is installed properly.
Consider the filter type (sulfur or phosphorus).
Check the system for leaks. (See “Checking for Leaks”.)
Verify that the method settings are appropriate.
Check the flow rates.
Check the makeup gas type.