EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
Peak Fronting
The figure below shows examples of the three types of peaks:
symmetric, fronting, and overloaded.
If peak fronting or overloading occurs, try the following:
Verify that the injection volume is appropriate. Column
overload is the most common cause for fronting peaks.
Decrease injection volume, dilute the sample, and/or
increase the split ratio.
Change the column type or film thickness.
Ensure that the column is installed properly. If not, reinstall
the column in the inlet.
Verify that the appropriate injection technique is being used.
This is usually related to erratic plunger depression or
having sample in the syringe needle.
If using capillary splitless injection, consider the compound
solubility in the injection solvent.
Change the solvent.
Use a retention gap.
Check purity of sample solvent. For solvents with large
differences in polarity or boiling points, a mixed sample
solvent can cause peak fronting. Change the sample solvent.
Symmetrical Fronting Overloaded

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals