88 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
TCD Peaks Have Negative Dip on Tail
Check for leaks at the detector column adapter fitting. (See
Checking for Leaks”.)
Upgrade the detector to a passivated filament.