EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
48 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
Peaks Not Displayed/No Peaks
If using an autosampler:
Ensure that there is sample in the vial.
Verify that the ALS plunger carriage is secured to the
syringe plunger.
Check that the syringe is installed correctly and draws
sample.
Verify that the turret/tray is loaded correctly and
injections are not from out-of-sequence vials.
Watch to see that the sample is pulled into the syringe.
Verify the detector in use is assigned to a signal.
Check the column for proper installation.
Ensure that the column is not plugged. (See “To Measure a
Column Flow”.) Perform column maintenance.
Check for leaks. (See “Checking for Leaks”.)
Check the flow settings, then measure the actual detector
flows. (See “To Measure a Detector Flow”.)
Some samples change with heat or ultraviolet light. Check
sample stability.
Check the sample level in the vial.
If the sample is viscous, try the following:
Increase the viscosity delay time.
Dilute the sample in an appropriate low-viscosity solvent.
Turn the tower fan off.
For 7683B ALS, use the tray quadrants to warm the
sample. For 7693A ALS, use the vial heater (accessory
G4514A Bar Code Reader/Mixer/Heater) to warm the
sample vial.
If the problem is with the detector, see Table 3.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals