EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
Troubleshooting Tasks 8
Agilent 7890 Series Troubleshooting 195
To Verify That the FID Flame Is Lit
To verify that the FID flame is lit, hold a mirror or other
reflective surface over the collector exhaust. Steady
condensation indicates that the flame is lit.
Typically the FID output will be between 5.0 and 20.0 pA when
lit and < 2.0 pA when not lit.
If the flame does not light, do the following:
Verify that the detector temperature is above 150 °C. Agilent
recommends operating the FID
300 °C.
Verify correct detector flows.
Inspect the jet for contamination.
Verify that the jet is installed correctly.
Check the column connections for leaks.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals