42 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
Retention Times Not Repeatable
Replace the septum.
Check for leaks in the inlet, liner (as applicable), and column
connection. (See “Checking for Leaks”.)
To Perform an Inlet Leak Check
To Check for Leaks in a Split/Splitless Inlet
To Perform a PP Pressure Decay Leak Test
To Check for Leaks in a Multimode Inlet
To Perform a COC Pressure Decay Leak Test
To Perform a PTV Pressure Decay Test
To Perform a VI Pressure Decay Test
Check for sufficient carrier gas supply pressure. The
pressure delivered to the GC must be at least 40 kPa (10 psi)
greater than the maximum inlet pressure required at final
oven temperature.
Run replicates of known standards to verify the problem.
Verify that you are using the correct liner type for the sample
being injected. (See Selecting the correct inlet liner.)
Consider if this is the first run. (Has the GC stabilized?)
If using an FID or NPD and retention times increase (drift),
check the jet for contamination or replace the jet.
To Replace an FID Jet
To Check for a Plugged FID Jet
To Maintain the NPD Collector, Ceramic Insulators, and
Jet
To Check for a Plugged NPD Jet

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals