ALS and Detector Symptoms 2
Agilent 7890 Series Troubleshooting 37
NPD Bead Will Not Ignite
Set the equilibration time to 0.0.
Verify that the flow settings are correct and appropriate. See
Flows, temperatures, and bead information.
If the process still fails, there could be a large leak in the
system. This causes measured flow rates to be different from
actual flow rates. Thoroughly leak check the whole system,
especially the detector column/adapter fitting. (See Checking
for Leaks.)
Check for fault messages. Press [Service Mode], then go to
Diagnostics > Front detector or Back detector > Detector signal.
You can also read the bead voltage.
Check the condition of the bead. Replace if necessary.
Inspect the jet to see if it is clogged. (See To Check for a
Plugged NPD Jet.)
Measure the actual detector flows. (See To Measure a
Detector Flow.) If the hydrogen or makeup flows are zero or
much lower than the displayed flow, suspect a plugged jet.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals