Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 43
Peak Areas Not Repeatable
Check the ALS syringe operation. (See the Troubleshooting
section of the sampler user manual: 7693A, 7650A, 7683B.)
Replace the syringe.
Check for leaks in the inlet, liner (as applicable), and column
connection. (See “Checking for Leaks”.)
Check sample level in vials.
Run replicates of known standards to verify the problem.
Consider if this is the first run. (Has the GC stabilized?)
For a multimode or split/splitless inlet in split mode, also check
for:
An abnormal split vent restriction. See:
To Run the Inlet Trap Check
To Perform a SS Split Vent Restriction Test.
To troubleshoot small leaks in a multimode or split/splitless
inlet, see To Check for Leaks in a Split/Splitless Inlet or To
Check for Leaks in a Multimode Inlet.