EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
Shutdown Symptoms 5
Agilent 7890 Series Troubleshooting 115
To recover from a hydrogen shutdown state:
1 Fix the cause of the shutdown:
Replace the inlet septum. See the Maintenance manual.
Replace the inlet O-ring. See the Maintenance manual.
Check for broken column.
Check the supply pressure. Make sure the gas supply
meets the pressure recommendations listed in the Agilent
GC, GC/MS, and ALS Site Preparation Guide.
Check the system for leaks. See Leak Check Tips.
2 Power cycle the GC.
3 After the GC powers back on, press the key for the device
that initiated the shutdown. Scroll to the pneumatic
parameter that is flashing Off, then press [On] or [Off]. For
example, if the front inlet ran out of carrier gas, press [Front
Inlet], scroll to the pressure or flow parameter, then press
[On].
After a column shutdown, you can turn on a detector gas
flow even if the keyboard is locked by a data system.
Hydrogen sensor
If equipped with the optional hydrogen sensor, the GC will
perform a hydrogen safety shutdown if the sensor measures an
uncombusted hydrogen concentration of about 1 % in the oven
(about 25 % of the lower explosive limit for hydrogen).

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals