EasyManua.ls Logo

Agilent Technologies 7890 Series - Page 134

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
134 Agilent 7890 Series Troubleshooting
7 Checking for Leaks
3
Verify that the Te s t pr essu re is acceptable. Typically, the
default pressure of 10 psi works well. If desired, enter a
different inlet pressure.
For repeatable results, use the same value for the same
hardware.
Enter a higher test pressure if using a column that
produces high backpressure.
4 Scroll to Test Inlet and press [On/Yes].
5 After a moment, the test stabilizes.
The TotalFlow reading shows the total flow of carrier gas
through the inlet. The Col reading shows the flow through
the column.
The approximate leak rate is TotalFlow – Col, in mL/min.
Consider the inlet leak free if the Col reading
approximately equals the TotalFlow reading.
6 While monitoring the readings, tighten fittings, replace the
septum, change the liner O-ring, and so forth as needed. If
the repair stops the leak, you will see the Col reading
decrease to be approximately equal to the TotalFlow reading.
7 If the test still fails:
Replace the septum
Reinstall the column into the inlet
Replace the liner and liner O-ring
Open the split vent trap and check the O-ring seating.
Replace the split vent trap if needed.
If the test passes, but you still suspect an inlet leak, perform a
pressure decay leak test. See:
To Perform a SS Pressure Decay Leak Test
To Perform a MMI Pressure Decay Test
To Perform a PP Pressure Decay Leak Test
To Perform a PTV Pressure Decay Test
To Perform a VI Pressure Decay Test
NOTE
If you ran the test on a leak-free inlet before performing maintenance, the
leak test result after maintenance should be about the same as it was
before maintenance.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Related product manuals