Checking for Leaks 7
Agilent 7890 Series Troubleshooting 139
1
Gather the following (see Consumables and parts for the
split/splitless inlet):
No-hole ferrule
1/4-inch wrench
Heat-resistant gloves (if the inlet is hot)
Column nut
New septum
O-ring
ECD/TCD Detector plug (part no. 5060-9055)
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Remove the column, if installed.
4 Plug the column fitting with a column nut and a no-hole
ferrule.
5 Remove the old septum and replace it with a new one. See To
change the septum on the split/splitless inlet.
6 Inspect the O-ring and replace it if it is hard and brittle or
cracked. See To change the liner and O-ring on the
split/splitless inlet.
7 Set the inlet to Split Mode.
The test can find leaks at the: The test cannot find leaks at the:
septum column fitting
septum nut gas supply bulkhead fittings to the
flow module
liner O-ring seal tubing and connections in a transfer
line connected to the inlet
gold seal/washer and reducing nut internal leaks in an EPC module
(septum purge valve)
inlet body
flow manifold split vent valve
flow manifold septum purge valve
split vent tubing and trap
septum purge tubing
seals within the tubing between the
inlet flow module and the inlet body

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals