EasyManua.ls Logo

Agilent Technologies 7890 Series - Page 142

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142 Agilent 7890 Series Troubleshooting
7 Checking for Leaks
For a GC with a G1888 Headspace Sampler:
A pressure drop of less than 2 psig in 5 minutes (0.4 psi/min
or less; 13.8 kPa or 2.76 kPa/min) is acceptable when using a
1 cc sample loop.
For all GCs:
Note that liner size impacts pressure drop. An inlet with a
smaller volume liner does not tolerate as large a leak rate as
an inlet with a larger volume liner.
21 After the inlet passes the test, restore the GC to operating
condition.
Restore source pressures.
Remove any caps/plugs.
Reinstall the column.
Restore the correct column configuration.
Load the operating method.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Related product manuals