EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
Checking for Leaks 7
Agilent 7890 Series Troubleshooting 169
changing the source pressure is not practical, skip this step.
If there is a relatively large volume in the supply tubing
leading up to the GC, consider whether or not there could
have been a leak into the system when evaluating the test
results.
18 Monitor the pressure for 10–15 minutes. Use the timer by
pressing [Time] and [Enter].
The pressure should drop approximately 1 psi (6.9 kPa)
during the first 1 to 2 minutes. After an initial pressure drop
of about 1 psi, the pressure should not drop more than
0.03 psi/min (0.21 kPa/min).
If the pressure drop is 0.03 psi/min or less, you can consider
the interface-gas sampler system leak-free.
If the pressure drops faster than the acceptable rate, you
must check the interface and sampler systems separately to
determine the source of the leak. See “To Prepare the VI for a
Closed System Leak Check” to create a closed flow system,
then return to this section and complete steps 10 to 16.
If you find a leak in the interface, refer to “To Correct Leaks
in the Volatiles Interface”.
If the interface is leak-free, pressure check the sampling
device. See the operating manual for your sampler for
instructions.
19 After the VI passes the test, restore the GC to operating
condition.
Restore source pressures.
Remove any caps/plugs.
If needed, reconnect the sampling device.
Reinstall the column.
Restore the correct column configuration.
Load the operating method.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals