EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 69
2
Measure the actual detector flow rates. See “To Measure a
Detector Flow”. If the actual hydrogen, makeup, and
capillary column flows are lower than the display, the jet is
becoming plugged. See “To Check for a Plugged FID Jet”.
Replace the jet.
3 Check that column is properly installed. Reinstall it. See To
Install a Capillary Column in the FID. Make sure the column
is installed all the way to the tip and withdrawn 1–2 mm
(column id > 100 μm).
4 Check inlet parameters that control venting, such as split
ratio and splitless purge delay time. Make sure the sample is
not being inadvertently vented.
5 Perform inlet maintenance (replace all consumable parts)
and pressure test the inlet when completed. See the
maintenance manual and “Checking for Leaks”.
6 Perform complete FID maintenance. Disassemble the FID
and clean or replace all parts. Refer to the following:
To Perform Maintenance on the FID Collector Assembly

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals