Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 71
Check that column is properly installed. Reinstall it. See To
Install a Capillary Column in the FID. Make sure the column
is installed all the way to the tip and withdrawn 1–2 mm
(column id > 100 μm).
Check for leaks at the detector column fitting. (See “Checking
for Leaks”.)