114 Advanced Operation Manual
3 Inlets
When is Start Run?
Both the inlet and oven temperature programs begin at Start
Run. All times—such as Purge time—are measured from Start
Run. When does Start Run occur?
• If the sample is injected manually, Start Run occurs when
the user presses the [Start] key.
• If a single injection per run is made using an
autosampler, Start Run occurs when the syringe carrier
moves down to make the injection.
• If multiple injections per run are made using an
autosampler, Start Run occurs when the syringe carrier
moves down to make the first injection of the set. There
are no Start Run signals for the rest of the injections in
the set.
Time Oven temperature Inlet temperature Inlet pressure Split vent flow
Between runs
Prep Run
Start Run
Vent end time
Purge time
Saver time
Rate 1
Initial time
Final temperature 1
Final time 1
Rate 1
Final temperature 1
Final time 1
Other rates,
temperatures,
and times, if
desired
Initial time
Other rates,
temperatures,
and times, if
desired
Saver or
Purge flow
Vent flow
Inlet is
pressure
controlled
Purge flow
Saver flow,
if on
Controlled by column
flow or pressure setpoint
or program
Vent pressure
Controlled by
column flow or
pressure setpoint
or program

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals