EasyManua.ls Logo

Agilent Technologies 7890 Series - Page 119

Agilent Technologies 7890 Series
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Inlets 3
Advanced Operation Manual 119
Set a Vent flow of 100 mL/min as a starting point.
Keep the inlet isothermal for now.
Enter a Vent pressure of 0 psi (0 kPa) and Vent end
time of 0.1 minutes.
d Make an injection using the empty vial. Use a
stopwatch or your GC's timer feature to time how long
the needle is in the inlet.
8 Enter revised Solvent vent mode parameters.
Set the method's Vent end time to be about 0.05 min
longer than the time the needle spends in the inlet.
Set the inlet temperature Initial time to be about 0.05
min longer than the Vent pressure time.
Program the inlet to ramp quickly to the injection
temperature. Make sure the ramp starts after the Vent
end time.
Set the Purge Flow to 30 mL/min. Set the purge time to
be the vent pressure Vent end time + 1 minute.
9 Make a 5 uL injection of your standard. You should see 5
times the response.
If the response of all analytes is too low:
The dispense speed is too fast. Liquid was injected into
the inlet and pushed out the vent.
The vent time is too long. The inlet started to heat
while the vent was open.
If the response of early eluters is too low:
The inlet temperature is too high.
The Vent flow is too high.
If the response of late eluters is too low:
The purge time is too short.
The final inlet temperature is too low.
10 If you need more injection volume and the 5 uL worked
to give 5 times the response, change the syringe to a
larger one, for example, 50 uL.
11 Set up the data system to perform a 25 uL injection.
12 Configure the syringe. Make sure the plunger speeds on
the injector are still set properly.
13 Recheck the injector timings. See the step above.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Related product manuals