Inlets 3
Advanced Operation Manual 121
Preinjection washes and pumps are performed only before
the first injection of a multiple injection set.
Postinjection washes are performed only after the last
injection in a multiple injection set.
An example
These values were used for a sample with a broad range of
boiling points.
Tabl e 24 General parameters
Name Value
Sample C
10
to C
44
hydrocarbons in hexane
Mode Solvent vent
MMI liner Glass wool packed
Injection volume One 10.0 μL injection (25 μL syringe)
Injection speed Fast
Column 30 m x 320 μm x 0.25 μm -5, part
number 19091J-413
Column flow 4 mL/min constant flow
Tabl e 25 Inlet parameters
Name Value Name Value
Initial temp 40 °C Rate 2 (off)
Initial time 0.3 min Pressure 15.6 psig
Rate 1 720 °C/min Vent pressure 0.0 psig
Final temp 1 450 °C Vent flow 100 mL/min
Final time 1 5 min Vent end time 0.2 min
Rate 2 100 °C/min Purge time 2.0 min
Final temp 2 250 °C Purge flow 50 mL/min
Final time 2 0 min
Tabl e 26 Oven parameters
Name Value
Initial temp 40 °C
Initial time 2.5 min

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals