Inlets 3
Advanced Operation Manual 135
Setpoint dependencies
Some setpoints in the flow system are interdependent. If you
change one setpoint, other setpoints may change to
compensate.
Initial values
The table shows recommended starting values for selected
parameters.
Tabl e 34 Setpoint dependencies
When you change These setpoints change
Column defined Column not defined
Purging You can change the
Pressure and Total flow
setpoints; other
setpoints are not
affected.
Purge flow Total flow**
Pressure Total flow**
Column flow*
Column flow
*
* This setpoint appears in the column parameters.
**This value is actual only.
Pressure
Total flow **
Before and after sampling, not purging You can change the
Pressure setpoint; other
setpoints are not
affected.
Pressure Column flow*
Total flow **
Column flow* Pressure
Total flow **
During sampling:
You cannot change pressure and flow setpoints during sampling time.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals