EasyManua.ls Logo

Agilent Technologies 7890 Series - Page 205

Agilent Technologies 7890 Series
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Detectors 5
Advanced Operation Manual 205
shown next to Makeup in the parameter list. Change
the gas type, if necessary.
b If your capillary column is defined, choose a flow
mode and set the makeup gas flow or combined flow.
c If your capillary column is not defined, enter a makeup
gas flow. Only constant flow is available.
7 Scroll to PMT voltage and make sure it is turned on.
8 Optionally, set the emission block temperature. (Normally,
the default temperature of 150 °C is sufficient for most
applications.) Scroll to Emission Block, input the desired
temperature, and press [Enter]. See “FPD+ temperature
considerations” on page 202 for more information.
9 Scroll to Flame and press [On/Yes]. This starts the
automatic ignition sequence (see “Lighting the FPD+
flame on page 203).
On successful ignition, the signal increases. Typical signal
output increases are 4 to 40 pA in sulfur mode, and 10 to
70 pA in phosphorus mode. You can also verify that the
flame is lit by holding a cold, shiny surface, such as a
mirror or chrome- plated wrench, over the vent exit.
Steady condensation indicates that the flame is lit.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Related product manuals