Flow and Pressure Modules 2
Advanced Operation Manual 29
When installing or replacing a frit, always use a new O- ring
(5180- 4181, 12/pk).
Selecting a frit
The frits change the control range of the channels. The
objective is to find a frit that allows the required range of
flows at reasonable source pressures.
For an auxiliary channel ordered as an option (part of
the GC order), use the frit supplied by the factory.
For an auxiliary channel ordered as an accessory
(separate from the GC order), see the instruction
information supplied with the accessory.
For a non- Agilent instrument, you must experiment to
find the appropriate frit.
When you change a frit, you change the physical
characteristics of the channel. It may be desirable (or
necessary) to change the PID constants for that channel. See
“PIDs” on page 21.