EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8753ET

Agilent Technologies 8753ET User Manual

Agilent Technologies 8753ET
479 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #152 background imageLoading...
Page #152 background image
2-14
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
Setting Measurement Parameters for the Power Meter Calibration
1. Connect the measurement equipment as shown in Step 1 of Figure 2-11.
Figure 2-11 Connections for R Channel and Source Calibration
CAUTION Note that the front panel jumper between R In and R Out must remain installed during the
procedure steps that use the connections in Step 1 of Figure 2-11.
2. From the front panel of the analyzer, set the desired receiver (RF) frequency and source output power by
pressing:
Note that these are the example RF start and stop frequencies. Enter the RF start and stop
frequencies for your measurement instead.
If the LO frequency is not set to 0 Hz, press:
3. To select the measurement trace, press:
The measurement trace is shown on the display.
4. Select the analyzer as the system controller:
System
INSTRUMENT MODE
FREQ OFFS MENU
650
M/
950
M/
LO MENU
FREQUENCY: CW
0
x1
Meas
INPUT PORTS
R
Local
SYSTEM CONTROLLER

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals