Chapter 6 203
Replacement Procedure
Synthesizer Module Replacement (A1)
6. Replacement Procedure
Synthesizer Module Replacement (A1)
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in) (for
reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)
Removal Procedure
Refer to Figure 6-21 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 167.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 168.
Step 3. Disconnect the semi-rigid cable (item 1).
Step 4. Disconnect the semi-rigid cable (item 2).
Step 5. Remove the two TORX T10 screws fastening the module to the bottom. (item 3)
Step 6. Slide the synthesizer module (item 4).