Chapter 7 249
Post-Repair Procedures
Post-Repair Procedures
7. Post-Repair Procedures
Receiver
Module (A7)
Perform the following required
adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic accuracy test
- Uncorrected system performance
Front End
Module (A9)
Perform the following required
adjustments
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Uncorrected System Performance Test
Doubler
Module (A6)
Perform the following required
adjustments.
- Source Output Power
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy Test
Table 7-2 Post-Repair Procedures for Analog Assemblies for Over 8.5 GHz Options