EasyManuals Logo

Agilent Technologies E5071C ENA User Manual

Agilent Technologies E5071C ENA
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Chapter 1 29
General Information
Power Meter Accuracy Test
1. General Information
NOTE It is recommended that a copy of the test record on page 32 be made, and the values be
recorded on the copy, thus preserving the original for future use.
Test Procedure
NOTE This procedure assumes the use of the recommended equipment model numbers. The
actual steps required, therefore, may differ for other model numbers of equipment used.
Step 1. Setup the equipment as shown in Figure 1-2.
a. Connect the DC REFERENCE OUTPUT connector on the rear panel of the range
calibrator to the DVM voltage input.
b. Connect the POWER METER output of the range calibrator to the input of the power
meter being tested.
c. Switch on the power to the power meter, the range calibrator, and the digital voltmeter.
Figure 1-2 Setup for the Power Meter Accuracy Test
Step 2. Press the power meter: Press [Preset/Local], then Confirm.
Step 3. Perform the following steps for each channel on the power meter:
a. Set to read in dBm: Press [dBm/W], then dBm.
b. Set the ref cal factor to 100%: Press [Zero/Cal], Cal, A/B Ref CF, then set to 100.0, if
necessary.
c. Set the cal factor to 100%: Press [Frequency/Cal Fac], A/B Cal Fac, then set to 100.0,
if necessary.
d. Set readout to 0.001 dBm: Press [Meas Setup], then Resolution 1234, to highlight 4.
e. Set filter step detect on and filter length to 512: Press [System/Inputs], channel Aor B
Input Setting
, [More], Ch A/B Filter, Step Det On, Filter On, Mode MAN, Length,
then set the filter length to
512.
Step 4. Setup the digital voltmeter (DVM) as follows:
a. Reset the DVM: Press the blue key followed by Reset.
b. Set the sample period to a value greater than one second: Press NPLC, 5, 0, then Enter.
Step 5. Set the range calibrator controls as follows:

Table of Contents