Chapter 4 97
Troubleshooting
Performance test failure troubleshooting
4. Troubleshooting
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
Receiver Port
Characteristics
SX
## ###
D1
# ## ###
D2
# ## ###
D3
# ## ###
D4
# ## ###
R1
## # ###
R2
## # ###
R3
## # ###
R4
## # ###
L21, L12
# ## ###
L31, L13
# ## ###
L41, L14
# ## ###
L32, L23
# ## ###
L42, L24
# ## ###
L43, L34
# ## ###
T21, T12
## ## # ## ###
T31, T13
## ## # ## ###
T41, T14
## ## # ## ###
T32, T23
## ## # ## ###
T42, T24
## ## # ## ###
T43, T34
## ## # ## ###
Table 4-12 Adjustment failure troubleshooting information (Over 8.5 GHz Options)
Probable faulty board assembly or parts
Failed
Adjustment
Failed test item
OCXO
BAS
SYN-SRC
SYN-LCL
DBL
RCV-1
RCV-2
RCV-3
RCV-4
DIS
IMX
FEM
CONN

Table of Contents