Programming Fundamentals 5
N9310A User’s Guide 99
IEEE 488.2 common command
Common commands are generally not measurement related,
but are used to manage macros, status registers,
synchronization, and data storage. The common commands
are defined by IEEE 488.2.
All common commands begin with an asterisk. The following
common commands are available in N9310A.
*RST (Reset)
Using this command resets the signal generator to factory
default settings and returns the instrument to a state where
it waits for a command to initiate other actions. See
“Preset” on page 70 for an overview of the factory default
settings of the signal generator.
*IDN? (Identification)
This is a query only command. After sending this query, the
signal generator reports the following identification
information:
Manufactory title
Instrument module
Instrument serial number
Firmware version number
*CLS (Clear Status)
Using this command clears the error queue of the signal
generator. There is no equivalent front panel key access.

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals