Subsystem Command Reference 6
N9310A User’s Guide 109
Amplitude Subsystem
The amplitude subsystem controls the settings and
parameters associated with amplitude.
Continuous Wave Output Amplitude
:AMPLitude:CW <val> <unit>
:AMPLitude:CW?
This command sets the amplitude of CW output.
The query returns the current amplitude of CW output.
Range and
unit
127 to +13 dBm, (+20 dBm settable)
80 to +60 dBmV, (+67 dBmV settable)
20 to +120 dBμV, (+127 dBμV settable)
0.0001 to 1000 mV, (2238.8 mV settable)
0.1 to 1 000 000 μV, (2238800 μV settable)
Minimum
increment
0.1 dB (When using dBm/dBmV/dBμV)
0.1 mV (When using mV)
0.1 μV (Ampl 999.9 μV)
100 μV (Ampl 1000 μV)
*RST value 127.0 dBm
Returned
data format
<val> <unit>
Front panel
access
Amplitude