Service Guide E8361-90001 3-45
PNA Series Microwave Network Analyzers Tests and Adjustments
E8361A Adjustments
Procedure
1. Connect the equipment as shown in Figure 3-17. Connect a GPIB cable between the
network analyzer and the power meter.
Figure 3-17 Setup 1 for the Receiver Calibration Adjustment
2. On the System menu, point to Service, Adjustments, and then click
Receiver Calibration.
3. Ensure the GPIB settings are correct.
4. Click Calibrate, and then follow the instructions as they are displayed.
Figure 3-18 Setup 2 for the Receiver Calibration Adjustment