EasyManuals Logo
Home>Agilent Technologies>Test Equipment>U1602B

Agilent Technologies U1602B User Manual

Agilent Technologies U1602B
227 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
7Performance Tests
Agilent U1602B/U1604B User’s and Service Guide 155
U1604B Frequency: 10 kHz 9.63 kHz to 10.4 kHz
Period: 100 ms 96.20 ms to 103.8 ms
Frequency: 1 MHz 962 kHz to 1.04 MHz
Period: 1 ms 0.96 ms to 1.04 ms
Frequency: 40 MHz 38.24 MHz to 41.93 MHz
Period: 25 ns 23.85 ns to 26.15 ns
Trigger Sensitivity
Calibrator Setting Te s t Lim i ts Channel 1 Channel 2
100 mVp-p at 5 MHz 0.8 division at 5 MHz [PASS/FAIL] [PASS/FAIL]
100 mVp-p at 20 MHz 1 division at 20 MHz (U1602B) [PASS/FAIL] [PASS/FAIL]
100 mVp-p at 40 MHz 1 division at 40 MHz (U1604B) [PASS/FAIL] [PASS/FAIL]
Multimeter Performance Verification Tests
DC Voltage Verification Test
Calibrator Voltage Range Test L i mit Results
600 mV 600 mV 597.7 mV to 602.3 mV
6 V 6 V 5.977 V to 6.023 V
30 V 60 V 29.86 V to 30.14 V
60 V 60 V 59.77 V to 60.23 V
600 V 600 V 597.7 V to 602.3 V
AC Voltage Verification Test
Calibrator Voltage Range Test L i mit Results
6 V at 1 kHz 6 V 5.928 V to 6.072 V
6 V at 30 kHz 6 V 5.808 V to 6.192 V
60 V at 1 kHz 60 V 59.28 V to 60.72 V
60 V at 30 kHz 60 V 58.08 V to 61.92 V
300 V at 50 Hz 600 V 295.80 V to 304.20 V
540 V at 1 kHz 600 V 533.40 V to 546.60 V
590 V at 10 kHz 600 V 571.1 V to 608.9 V

Table of Contents

Other manuals for Agilent Technologies U1602B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1602B and is the answer not in the manual?

Agilent Technologies U1602B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1602B
CategoryTest Equipment
LanguageEnglish

Related product manuals