7Performance Tests
Agilent U1602B/U1604B User’s and Service Guide 155
U1604B Frequency: 10 kHz 9.63 kHz to 10.4 kHz
Period: 100 ms 96.20 ms to 103.8 ms
Frequency: 1 MHz 962 kHz to 1.04 MHz
Period: 1 ms 0.96 ms to 1.04 ms
Frequency: 40 MHz 38.24 MHz to 41.93 MHz
Period: 25 ns 23.85 ns to 26.15 ns
Trigger Sensitivity
Calibrator Setting Te s t Lim i ts Channel 1 Channel 2
100 mVp-p at 5 MHz 0.8 division at 5 MHz [PASS/FAIL] [PASS/FAIL]
100 mVp-p at 20 MHz 1 division at 20 MHz (U1602B) [PASS/FAIL] [PASS/FAIL]
100 mVp-p at 40 MHz 1 division at 40 MHz (U1604B) [PASS/FAIL] [PASS/FAIL]
Multimeter Performance Verification Tests
DC Voltage Verification Test
Calibrator Voltage Range Test L i mit Results
600 mV 600 mV 597.7 mV to 602.3 mV
6 V 6 V 5.977 V to 6.023 V
30 V 60 V 29.86 V to 30.14 V
60 V 60 V 59.77 V to 60.23 V
600 V 600 V 597.7 V to 602.3 V
AC Voltage Verification Test
Calibrator Voltage Range Test L i mit Results
6 V at 1 kHz 6 V 5.928 V to 6.072 V
6 V at 30 kHz 6 V 5.808 V to 6.192 V
60 V at 1 kHz 60 V 59.28 V to 60.72 V
60 V at 30 kHz 60 V 58.08 V to 61.92 V
300 V at 50 Hz 600 V 295.80 V to 304.20 V
540 V at 1 kHz 600 V 533.40 V to 546.60 V
590 V at 10 kHz 600 V 571.1 V to 608.9 V