Performance Tests 7
156 Agilent U1602B/U1604B User’s and Service Guide
Resistance Verification Test (2-Wire Ohms)
Calibrator Settings Range Tes t L imit Results
600 W 600 W 595.8 W to 604.2 W
6 kW 6 kW 5.958 kW to 6.042 kW
60 kW 60 kW 59.58 kW to 60.42 kW
600 kW 600 kW 595.8 kW to 604.2 kW
3 MW 6 MW 2.973 MW to 3.027 MW
6 MW 6 MW 5.958 MW to 6.042 MW
60 MW 60 MW 59.28 MW to 60.72 MW
Capacitance Verification Test
Calibrator Settings Range Tes t L imit Results
60 nF 60 nF 58.68 nF to 61.32 nF
600 nF 600 nF 586.8 nF to 613.2 nF
6 mF6 mF 5.868 mF to 6.132 mF
36 mF 60 mF 35.16 mF to 36.84 mF
60 mF 60 mF 58.68 mF to 61.32 mF
300 mF 300 mF 293.4 mF to 306.6 mF
Diode Verification Test
Calibrator Voltage Tes t Limi t Results
0.9 V 0.8812 V to 0.9188 V
Continuity Verification Test
Calibrator Settings Condition Results
30 W Buzzer sounds on [PASS/FAIL]
75 W Buzzer sounds off [PASS/FAIL]