Sequoia Series    119 
VOLTage:SLEW 
Phase Selectable 
This command sets the slew rate for all programmed changes in the output voltage level of the source. 
A parameter of MAXimum will set the slew to its maximum possible rate. The SCPI representation for 
MAXimum is 1E9. 
Command Syntax  [SOURce:]VOLTage:SLEW[:IMMediate]<NRf+> 
Parameters  0 to1E9 | MAX  
Unit  V/S (volts per second) 
*RST Value  MAX 
Examples  VOLT:SLEW 1  VOLT:SLEW MAX 
Query Syntax  [SOURce:]VOLTage:SLEW[:IMMediate]? 
Returned Parameters  <NR3>  
Related Commands  VOLT:SLEW:MODE  VOLT:SLEW:TRIG 
VOLTage:SLEW:MODE 
Phase Selectable 
This command determines how the output voltage slew rate is controlled during a triggered output 
transient. The choices are:  
FIXed  The slew rate is unaffected by a triggered output transient. 
STEP  The slew rate is programmed to the value set by VOLTage:SLEW:TRIGgered 
when a triggered transient occurs. 
PULSe  The slew rate is changed to the value set by VOLTage:SLEW:TRIGgered for 
a duration determined by the pulse commands. 
LIST  The slew rate is controlled by the voltage slew list when a triggered transient 
occurs. 
Command Syntax  [SOURce:]VOLTage:SLEW:MODE<mode> 
Parameters  FIXed | STEP | PULSe | LIST 
*RST Value  FIX 
Examples  VOLT:SLEW:MODE LIST  VOLT:SLEW:MODE FIX 
Query Syntax  [SOURce:]VOLTage:SLEW:MODE? 
Returned Parameters  <CRD> 
Related Commands  VOLT:SLEW:TRG  VOLT:SLEW 
VOLTage:SLEW:TRIGgered 
Phase Selectable 
This command  selects  the slew rate  that  will be set  during a triggered  step or  pulse transient. A 
parameter of MAXimum will set the slew to its maximum possible rate. The maximum allowed value 
is1E9. 
Command Syntax  [SOURce:]VOLTage:SLEW:TRIGgered<NRf+> 
Parameters  0 to 1E9 | MAXimum 
Unit  V/S (volts per second) 
*RST Value  MAXimum 
Examples  VOLT:SLEW:TRIG 1  VOLT:SLEW:TRIG MAX 
Query Syntax  [SOURce:]VOLTage:SLEW:TRIGgered? 
Returned Parameters  <NR2>  
Related Commands  VOLT:SLEW:MODE  VOLT:SLEW