140 Sequoia Series
INITiate:CONTinuous
These commands control the transient generator trigger system.
• 1 or ON continuously initiates the transient trigger system.
• 0 or OFF turns off continuous triggering. In this state, the trigger system must be initiated
for each triggered event using INITiate:TRANsient.
Command Syntax INITiate:CONTinuous
Parameters 0|1|OFF|ON
*RST Value OFF
Examples INIT:CONT ON
Related Commands ABOR INIT:CONT *TRG
TRIGger[:TRANSient]:SOURce
This command will determine the source of the trigger event for the transient. The following sources
are available:
• IMMediate Triggering occurs as soon as the INIT command is received.
• BUS Triggering occurs following the INIT command after receiving the *TRG
command or a Group Execute Trigger (GET) IEEE signal.
• EXTernal Triggering occurs when an external trigger input is received.
Command Syntax TRIGger[:TRANSient]:SOURce
Parameters IMMediate | BUS | EXT
*RST Value IMMediate
Examples TRIG:SOUR BUS
Related Commands ABOR <group execute trigger>
TRIGger:COUNt
This command determines whether the phase synchronization is accomplished at the beginning of the
transient list or pulse for every COUNt or at the begining of the transient for the first count only. It
accepts the following parameters.
• NONe Synchronization is accomplished once
• ALL Synchronization is accomplished for every count.
Command Syntax TRIGger:COUNt ALL
TRIGger:COUNt NONE
Parameters ALL, NONE
*RST Value NONE
Examples TRIG:COUN ALL
TRIG:COUN NONE
Query Syntax TRIG:COUN?
Returned Parameters <CRD>
Related Commands ABOR TRIG TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG WAI