TRIGger:ACQuire:SOURce   
These commands select the trigger source for a triggered measurement sequence as follows: 
IMMediate  Acquire  the  measurements  immediate.  The  FETCh  commands  return  the 
requested  calculation  from  the  set  of  acquired  data  as  soon  as  the 
measurement calculations are completed by the controller. 
SYNChronize  Internal phase A angle 
BUS  IEEE-488 device, *TRG, or <GET> (Group Execute Trigger) 
TTLTrg  The signal driving the Trigger Out BNC 
Command Syntax  TRIGger:ACQuire:SOURce<CRD> 
Parameters  IMM|SYNC|BUS |TTLT 
*RST Value  IMM 
Examples  TRIG:ACQ:SOUR BUS 
Query Syntax  TRIGger:ACQuire:SOURce? 
Returned Parameters  <CRD> 
Related Commands  ABOR  TRIG:SOUR  TRIG:SYNC:SOUR 
  TRIG:SYNC:PHAS  INIT  INIT:CONT  *TRG   
TRIGger:SYNChronize:SOURce 
This command selects the synchronizing trigger source in generating a step, pulse, or list output as 
follows: 
•  IMMediate  starts the transient output immediately. 
•  PHASe  starts the transient output at the reference phase set by TRIG:PHAS. 
Command Syntax  TRIGger:SYNChronize:SOURce <CRD> 
Parameters  IMMediate|PHASe 
*RST Value  IMM 
Examples  TRIG:SYNC:SOUR IMM  TRIG:SOUR PHAS 
Query Syntax  TRIGger:SYNC:SOURce? 
Returned Parameters  <CRD> 
Related Commands  ABOR   
  TRIG:SYNC:PHAS  INIT  INIT:CONT  *TRG 
TRIGger:SYNChronize:PHASe 
This command sets the phase angle with respect to an internal phase reference if TRIGger:SOURce 
PHASe is selected. The range is from -360 to +360 Degrees. 
Command Syntax  TRIGger:SYNChronize:PHASe<NRf+> 
Parameters  360 to +360 
*RST Value  0 
Examples  TRIG:SYNC:PHAS 90   
Query Syntax  TRIGger:SYNChronize:PHASe? 
Returned Parameters  <NR2> 
Related Commands  ABOR  TRIG:SYNC:SOUR  INIT 
INIT:CONT  *TRG