Sequoia Series 195
IEC413:SWEep:LEVel
This command will set the level for the harmonics sweep test. The command will accept either one
parameter or five parameters. In case of one parameter, the level will be set for the selected range
only. The IEC413:STATe must be in the off position and the user class is selected to accept the
command.
Command Syntax IEC413:SWEep:LEVel <NRf>[,<NRf1.. NRf4]>
Parameters 0 to 20%
Examples IEC413:SWE:LEVel 10
Query Syntax IEC413:SWE:LEVel?
Returned Parameters <NR2>
IEC413:SWEep:DWELl
This command will set the dwell time in seconds for the entire sweep range. The class must be selected
prior to this command. Query with min and max to find the range of dwell time.
Command Syntax IEC413:SWEep:DWELl <NRf+>
Parameters <NRf>
Examples IEC413:SWE:DWEL 200
Query Syntax IEC413:SWE:DWEL?
Returned Parameters <NR1>
IEC413:SWEep:CURRent[:DATA]?
This command will return the measured harmonic current. This data is available after the sweep is
completed.
Command Syntax IEC413:SWEep:CURRent[:DATA] ]?
Parameters none
Query Syntax IEC413:SWE:CURR?
Returned Parameters <NR2>,.... ,<NRn>
IEC413:SWEep:FREQuency[:DATA]?
This command will return the frequency data points that correspond to the harmonic current data
points. This data is available after the sweep is completed.
Command Syntax IEC413:SWEep:FREQuency?
Parameters none
Query Syntax IEC413:SWE:CURR:FREQ?
Returned Parameters <NR2>,.... ,<NRn>