Sequoia Series 197
IEC413:HARMonics:TABle
This command allows selection of only one of the Individual harmonics tables of the IEC 61000-4-13
standard. Default is to run all tables but some EUT’s for class 3 may require only Table 1 to be run.
Available selections are Table 1, 2, 3 or all tables.
Command Syntax IEC413:HARMonics:TABle <CRD>
Parameters 1 | 2 | 3 | ALL
Examples IEC413:HARM:TAB ALL
Query Syntax IEC413:HARM:TAB?
Returned Parameters <CRD>
IEC413:HARMonics:DWELl
This command will set the dwell time in seconds for each harmonic. The class must be selected prior
to this command. Query with min and max to find the range of the dwell time.
Command Syntax IEC413:HARM:DWELl <NRf+>
Parameters <seconds>
Examples IEC413:HARM:DWEL 10
Query Syntax IEC413:HARM:DWEL?
Returned Parameters <NR1>
IEC413:HARMonics:PAUSe
This command will set the pause time in seconds. This is the time between each harmonic. The class
must be selected prior to this command. Query with min and max to find the range of the dwell time.
Command Syntax IEC413:HARM:PAUSe <NRf+>
Parameters <seconds>
Examples IEC413:HARM:PAUS 2
Query Syntax IEC413:HARM:PAUS?
Returned Parameters <NR1>
IEC413:IHARMonics:LEVel
This command will set the level for the interharmonics test. The command will accept either one
parameter or five parameters. In case of one parameter, the level will be set for the selected range
only. The IEC413:STATe must be in the off position and the user class must be selected to accept
the command.
Command Syntax IEC413:IHARmonics:LEVel <NRf>[,<NRf1.. NRf4]>
Parameters 0 to 20%
Examples IEC413:IHAR:LEV 10
Query Syntax IEC413:IHAR:LEV?
Returned Parameters <NR2>