38 Sequoia Series
4.2.5 Voltage Measurement
CALibration:MEASure:VOLTage[:AC][:FSCale] <NRf>
This command initiates the calibration of the RMS voltage measurement at full scale.
Command Syntax CALibration:MEASure:VOLTage[:AC][:FSCale]
Parameters <NRf> (actual rms output voltage measured with external device)
Examples CAL:MEAS:VOLT 120
Query Syntax CALibration:MEASure:VOLTage?
Returned Parameters <NR2> (value range -1000 to +1000)
CALibration:MEASure:VOLTage:DC[:FSCale] <NRf>
This command initiates the calibration of the DC voltage measurement at full scale.
Command Syntax CALibration:MEASure:VOLTage:DC[:FSCale]
Parameters <NRf> (actual DC output voltage measured with external device)
Examples CAL:MEAS:VOLT:DC 120
Query Syntax CALibration:MEASure:VOLTage:DC?
Returned Parameters <NR2> (value range -1000 to +1000)
CALibration:MEASure:VOLTage:DC:NEGative <NRf>
This command initiates the calibration of the negative DC voltage measurement at full scale.
Command Syntax CALibration:MEASure:VOLTage:DC:NEGative
Parameters <NRf> (actual DC output voltage measured with external device)
Examples CAL:MEAS:VOLT:DC -120
Query Syntax CALibration:MEASure:VOLTage:DC:NEG?
Returned Parameters <NR2> (value range -1000 to +1000)