Sequoia Series 37
This command initiates the calibration of the DC current measurement at full scale.
Command Syntax CALibration:MEASure:CURRent:DC[:FSCale]
Parameters <NRf> (actual load current measured with external device)
Examples CAL:MEAS:CURR:DC 11.5
Query Syntax CALibration:MEASure:CURRent:DC?
Returned Parameters <NR2> (value range -1000 to +1000)
CALibration:MEASure:CURRent:DC:LROFset?
This command will set the low range DC current measurements offset value
Command Syntax CALibration:MEASure:CURRent:DC:LROFset
Parameters <NRf> (actual value reported with the command MEAS:CURR:DC?
Examples CAL:MEAS:CURR:DC:LROF 0.020
Query Syntax CALibration:MEASure:CURRent:DC:LROFset?
Returned Parameters <NR2>
CALibration:MEASure:CURRent:DC:HROFset?
This command will set the high range DC current measurements offset value
Command Syntax CALibration:MEASure:CURRent:DC:HROFset
Parameters <NRf> (actual value reported with the command MEAS:CURR:DC?
Examples CAL:MEAS:CURR:DC:HROF 0.020
Query Syntax CALibration:MEASure:CURRent:DC:HROFset?
Returned Parameters <NR2>
CALibration:MEASure:CURRent:DC:NEGative <NRf>
This command initiates the calibration of the negative DC current measurement at full scale.
Command Syntax CALibration:MEASure:CURRent:DC:NEGative
Parameters <NRf> (actual load current measured with external device)
Examples CAL:MEAS:CURR:DC -11.5
Query Syntax CALibration:MEASure:CURRent:DC:NEG?
Returned Parameters <NR2> (value range -1000 to +1000)