36 Sequoia Series
Command Syntax CALibration:IHARmonic?
Related Commands CAL:VOLT:HFR
4.2.3 Save
CALibration:SAVE
This command saves all calibration coefficients to non-volatile memory. This command should be
issued after all calibration adjustments have been made. If not, all changes will be lost when unit is
turned off and the previous calibration values will take effect the next time the unit is powered up.
Note: Saving calibration data to non-volatile memory requires more time to process by the
Sequoia controller than other commands. As such, it is recommended to hold off on
sending additional commands for about 300 msecs
Command Syntax CALibration:SAVE<SRD>
Parameters None
Examples CAL:SAVE
Related Commands CAL:PASS
4.2.4 Current Measurement
CALibration:MEASure:CURRent[:AC][:FSCale] <NRf>
This command initiates the calibration of the AC current measurement at full scale.
Command Syntax CALibration:MEASure:CURRent[:AC][:FSCale]
Parameters <NRf> (actual load current measured with external device)
Examples CAL:MEAS:CURR 11.5
Query Syntax CALibration:MEASure:CURRent?
Returned Parameters <NR2> (value range -1000 to +1000)
CALibration:MEASure:CURRent[:AC]:LROFset?
This command will set the low range current measurements offset value
Command Syntax CALibration:MEASure:CURRent[:AC]:LROFset
Parameters <NRf> (actual value reported with the command MEAS:CURR?
Examples CAL:MEAS:CURR[:AC]:LROF 0.020
Query Syntax CALibration:MEASure:CURRent[:AC]:LROFset?
Returned Parameters <NR2>
CALibration:MEASure:CURRent[:AC]:HROFset?
This command will set the high range current measurements offset value
Command Syntax CALibration:MEASure:CURRent[:AC]:HROFset
Parameters <NRf> (actual value reported with the command MEAS:CURR?
Examples CAL:MEAS:CURR[:AC]:HROF 0.020
Query Syntax CALibration:MEASure:CURRent[:AC]:HROFset?
Returned Parameters <NR2>
CALibration:MEASure:CURRent:DC[:FSCale] <NRf>