Sequoia Series    35 
  [:AC] 
    :LROFset    Calibrate the offset voltage at the low range 
    :HROFset    Calibrate the offset voltage at the high range 
  :DC 
    :LROFset    Calibrate the offset voltage at the low range 
    :HROFset    Calibrate the offset voltage at the low range 
:CURRent 
  [:AC] 
    :LROFset    Calibrate the offset current at the low range 
    :HROFset    Calibrate the offset current at the low range 
  :DC 
    :LROFset    Calibrate the offset current at the low range 
    :HROFset    Calibrate the offset current at the low range 
  :MONitor 
:VOLTage 
  [:FSCale]      Calibrate the coefficient for Voltage Monitor VMON 
  :ZERO        Cancel voltage monitor offset 
:CURRENT 
  [:FSCale]      Calibrate the coefficient for Current Monitor IMON 
  :ZERO        Cancel current monitor offset 
   
4.2.1  Password 
CALibration:PASSword 
This command allows the entry of the calibration password. The calibration password is required to 
use the data entry form of the calibration commands. Without the use of this password, only the query 
form  can  be  used  to  query  any  calibration  coefficient,  but  no  new  calibration  can  be  performed. 
Calibration queries always return two values. The first value is the calibration coefficient  itself; the 
second value is the temperature associated with that coefficient. All temperate coefficients except for 
full-scale AC voltage are computed by the power source controller. 
The calibration password is defined as the numeric portion of the power source serial number spelled 
backwards. The password needs to be enclosed by single or double quotation marks. Thus, if the 
unit’s serial number is HK12345, the calibration password is “54321” and the command syntax would 
be: 
CAL:PASS “54321” 
Note that any non-numeric characters such as the HK in the example shown here need to be discarded 
when sending the calibration password. Only the numeric portion is to be used. 
Command Syntax  CALibration:PASSword<SRD> 
Parameters  <numeric portion of serial number reversed> (default) 
Examples  CAL:PASS '34593'  CAL:PASS "35461" 
Related Commands  *IDN? 
4.2.2  IHARmonic 
CALibrate:IHARmonic 
This query command retrieves the interharmonic calibration coefficient for the IEC413 option. 
This query returns a comma-separated list of the calibration coefficient and the frequency at 
which the calibration was peformed.