Sequoia Series    87 
REGenerate:DELay:OUTPut[:SHUTdown]  
This command will set the delay in seconds before  shutting down the power source output in an 
attempt to shut down the inverter under test. Available setting range is 0.25 to 10.0 seconds. If the 
frequency deviation is set to 0.0 Hz, no frequency shift will be asserted. If set to a value other than 
zero hertz, the frequency shift will be applied first when the fed back current exceeds the programmed 
current limit for more than the delay set by the REG:DEL:FREQ command. If the current fed back by 
the EUT still exceeds the programmed current limit, the output voltage will be set to the lower voltage 
limit and the output relay will open after this delay expires. 
Command Syntax  REGenerative:DELay:OUTPut [:SHUTdown] <NRf+> 
Parameters  0.25 to 10, time in seconds 
Examples  REGenerative:DELay:OUTPut [:SHUTdown] 10 
Query Syntax  REGenerative:DELay:OUTPut [:SHUTdown]? 
Returned Parameters  <NRf+> 
Related Commands  REG:DEL:OUTP?MIN, REG:DEL:OUTP?MAX  
REGenerate:CURRent 
This command will set or query the regenerate current limit over which the EUT is allowed to operate. 
Note that this current limit set point is different from the normal mode current limit setting. 
Command Syntax  REGenerative:CURRent <NRf+> 
Parameters  0 to max. value for selected current range 
Unit  A (rms amperes) 
Examples  REGenerative:CURRent 10 
Query Syntax  REGenerative:CURRent? 
Returned Parameters  <NRf+> 
Related Commands  REG:CURR?MIN, REG:CURR?MAX