EasyManua.ls Logo

AMI 1700 - Figure 4-27 MENU Selection Button; Figure 4-28 SENSORS Selection Button; Figure 4-29 CALIBRATE NITROGEN Selection Button; Figure 4-30 PERFORM MAX CAL. Selection Button

AMI 1700
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50 Rev. 2
Calibration
Capacitance-Based Level Calibration : Approximate Calibration Procedure
sensor cannot be dipped, then a partial length dip can be
performed. If both situations are encountered, then a partial length
dip can be performed in a substitute reference liquid.
1. Press the MENU button in the
lower left corner of the display
screen.
2. Choose the SENSORS
selection from the MENU
screen.
3. Press the CALIBRATE
NITROGEN button
4. Position the capacitance
sensor in the target liquid at
the 100% level. Hold the
sensor at this location and
press the PERFORM MAX
CAL button.
Figure 4-27. MENU Selection Button
Figure 4-28. SENSORS Selection
Figure 4-29. CALIBRATE NITROGEN
Selection Button
Figure 4-30. PERFORM MAX CAL.
Selection Button

Table of Contents

Related product manuals