EasyManua.ls Logo

Arkray ADAMS A1C HA-8180V - Page 100

Arkray ADAMS A1C HA-8180V
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HA-8180V 2-43
c Below <Calibration method>, select <Calibration coefficients
setting>.
Hyphen button: Changes the option.
button: Moves the cursor to the next setup field.
d Press .
2 Set the calibration coefficients.
a Enter the calibration coefficients (“a” and “b”).
See the table below.
button: Moves the cursor to the next setup field.
b Press .
The message “Settings changed. Save setting changes?” will
appear.
c Press .
This saves your entry and will return you to the standby screen.
HbA1c calibration
10-06-2010 10:00
Standard value (mmol/mol)
L < 0> H < 0>
Conversion to NGSP: Y = a X + b
a: <0.0915>
b: <+2.15>
Calibration method
[Calibration coefficients setting ]
PrevItem OK Cancel
OK
HbA1c calibration
10-06-2010 10:00
HbA1c (mmol/mol) Y = a X + b
a: <1.0000>
b: <+ 0.0>
Conversion to NGSP: Y = a X + b
a: <0.0915>
b: <+2.15>
PrevItem BS OK Cancel
OK
OK
Setup item Description
HbA1c (mmol/mol) Enter <a> as the gradient and <b> as the intercept of the calibration coefficients.
a: 0.0000 to 1.5000 (Default: 1.0000), b: -50.0 to 50.0 (Default: 0.0)
Conversion to NGSP Enter coefficients “a” and “b” of the conversion formula, “Y=aX+b”, to convert HbA1
C values
from the IFCC unit (mmol/mol) to the NGSP unit (%).
a: 0.0000 to 0.1500, b: -5.00 to 5.00
* The default is the value that appears below <Conversion to NGSP> on the screen in step
1-c.
2.8 HbA1C Calibration

Table of Contents

Related product manuals