8. Definitions and parameters
User Manual – PhysioGo.Lite Electro page 70 / 96
Symbol Description
the treatment
Available parameters
Pause duration t_pause
Non-editable parameter, calculated from the formula:
• ASYM: (1/f_bas)-t_imp
•
Basic pulse
frequency
f_bas
Regulation in the range of 0,1 Hz – 500 Hz (variable step).
Remark: the maximum frequency value depends on the
pulse duration, see Table 8.2.
Amplitude
modulation
program
PT
Amplitude modulation program is switched
off
Contraction phase 10 s
Fall time 2 s
Polarization
NOR
For such polarization setting red plug is a
positive electrode, and black plug is a
REV
For such polarization setting red plug is a
negative electrode and black plug is a
positive electrode. Warning means that such
a setting is the reverse of commonly
accepted way of marking polarity.
ALT
Automatic polarization switch in the half
performed treatment.
operation
-
CC – stabilized output current
CV – stabilized output voltage
Maximum
amplitude
-
Unipolar
current
Regulation step:
• 0,1 mA in the range of 0,3-10 mA
• 0,5 mA in the range of 10-140 mA
1,5 – 100 V in CV mode, max. 80 mA
Regulation step:
•
0,5 V in the range of 1,5-100 V
Bipolar
current
Regulation step:
• 0,1 mA in the range of 0,5-10 mA
• 0,5 mA in the range of 10-140 mA
1,5 – 100 V in CV mode, max. 80 mA
Regulation step:
• 0,5 V in the range of 1,5-100 V
Table 8.2. Available frequency values in relation to pulse duration
[ms]
basic pulse
pulse frequency
basic pulse
frequency possible to