Page 34 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Bolus in mL/hr and mL/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default
bolus size may appear automatically. If a default bolus size is displayed, use
and or the digit keys to modify the value as necessary.
If the infusion program is to include a bolus delivery:
1. Press BOLUS to obtain the bolus programming display.
2. Press EDIT, then use the digit keys to set the bolus size (in mL).
3. Press CONFIRM to store the bolus size and return to the normal display. The bolus
infusion rate is automatically set.
See “Bolus Function” on page 55 for more information on bolus delivery.
Note: For mL/min mode, the bolus rate is displayed in mL/min. For all other modes, it is
displayed in mL/hr.
mL/hr and mL/min Programming Notes
1. See “Syringe Data Tables” on page 25 for minimum and maximum infusion rates for
each syringe.
2. Pressing CONFIRM during delivery briefly displays additional program data.
3. A bolus can only be programmed, edited, or delivered after the syringe size has been
programmed.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals