07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 39
Detailed Instructions
Programming CUSTOM DILUTION Mode
In this mode, the pump runs at a programmed, constant rate. The infusion is programmed
in terms of target rate, patient body weight, initial drug concentration, target dose, and
final volume. The pump calculates and displays the drug volume to be mixed with
sufficient IV base fluid to yield the programmed final volume. The final concentration is
displayed when a bolus or infusion is initiated.
Note: The rate, body weight, concentration, and dose fields will not be editable after the
confirm key is pressed. Power the pump off and then back on to make these fields
editable again.
1. At the SELECT MODE ▲▼ prompt, use or to display the available infusion mode
selections. Press
CONFIRM to select the CUSTOM DILUTION mode.
2. Use or and CONFIRM as required to select the syringe manufacturer and size.
3. Enter the target rate (RATE) in mL/hr. Press CONFIRM to complete the data entry.
4. Enter the patient body weight (BODY WT) in kg. Press CONFIRM to complete the data
entry.
5. Enter the initial drug concentration (CONC.) in mg/mL. Press CONFIRM to complete
the data entry.
6. Enter the target dose (DOSE) in mcg/kg/min. Press CONFIRM to complete the data
entry.
7. Enter the final volume (FINAL VOL) in mL. Press CONFIRM to complete the data entry.
8. In the DRUG VOL field, the pump displays the volume of drug that must be mixed with
IV base fluid to yield the programmed final volume. The text field displays
DRUG VOL.
Press
CONFIRM to acknowledge the drug volume.
9. Mix the indicated drug volume (DRUG VOL) with sufficient IV base fluid to produce the
final volume that was programmed (
FINAL VOL).
Note: The solution may be mixed directly in the syringe, or may be mixed in an IV bag
and then drawn up into the syringe.
10. Attach a 2N Series (or equivalent) IV extension set to the syringe and manually purge
the air from the filled syringe and tubing.
11. Mount the syringe.
12. Perform a purge operation to remove slack from the plunger driver.
13. After completing the purge procedures, press START. The new final concentration will
be displayed in the concentration field.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals