Additional Functions and Operations
Page 54 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
purge accumulated volume will disappear when the START key is pressed and will not
be included in the total volume.
4. Repeat steps 2 and 3 as often as necessary to obtain a steady drip at the end of the
tubing.
Purge Delivery Table
Syringe
Manufacturer
Syringe Size
(mL)
Approximate
Purge
Volume (mL)
Purge Rate
(mL/hr)
Purge Rate
(mL/min)
B-D Plastic 1
3
5
10
20
30
60
0.02
0.08
0.15
0.22
0.39
0.50
0.75
11
38
74
107
188
241
366
0.19
0.63
1.2
1.7
3.1
4.0
6.1
Monoject 1
3
6
12
20
35
60
0.02
0.09
0.17
0.26
0.43
0.59
0.75
11
41
83
126
208
285
366
0.19
0.68
1.3
2.1
3.4
4.7
6.1
Teru mo 1
3
5
10
20
30
60
0.02
0.09
0.18
0.27
0.44
0.58
0.75
11
41
87
129
210
276
438
0.19
0.69
1.4
2.1
3.5
4.6
7.3
NOTE: Purge volume and rate information shown applies to pumps configured for Rate Range = H (High).
NOTE: Purge volumes are approximately the amount delivered in a single complete purge delivery, without
slack in the drive system.
NOTE: The volume of purged fluid is not added to the amount in the TOTAL mL field.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals